Improved real‐time tagged MRI using REALTAG
Abstract
Objectives
To evaluate a novel method for real‐time tagged MRI with increased tag persistence using phase sensitive tagging (REALTAG), demonstrated for speech imaging.
Methods
Tagging is applied as a brief interruption to a continuous real‐time spiral acquisition. REALTAG is implemented using a total tagging flip angle of 180° and a novel frame‐by‐frame phase sensitive reconstruction to remove smooth background phase while preserving the sign of the tag lines. Tag contrast‐to‐noise ratio of REALTAG and conventional tagging (total flip angle of 90°) is simulated and evaluated in vivo. The ability to extend tag persistence is tested during the production of vowel‐to‐vowel transitions by American English speakers.
Results
REALTAG resulted in a doubling of contrast‐to‐noise ratio at each time point and increased tag persistence by more than 1.9‐fold. The tag persistence was 1150 ms with contrast‐to …
- Date
- January 1, 1970
- Authors
- Weiyi Chen, Nam Gyun Lee, Dani Byrd, Shrikanth Narayanan, Krishna S Nayak
- Journal
- Magnetic resonance in medicine
- Volume
- 84
- Issue
- 2
- Pages
- 838-846